100 SLPM Flow
Sample make/model: ALICAT · MBS-100SLPM-D
Description
Proficiency test for flow meter calibration at 100 SLPM. Demonstrate competence with your laboratory flow procedures and measurement uncertainty.
Sample instructions
To complete this PT you will need a flow standard set to mass flow in slpm and an air gas supply. When you receive the test flow meter marked FLOW-115 please allow 24 hours before the test to acclimate to room temperature.
- Connect the flow meter marked FLOW-115 to your flow standard and air supply.
- Set flow conditions on your flow standard to 1 atm and 21.11 °C.
- Set flow to 25 slpm and record reading. Report reading and uncertainty on the Data Reporting Sheet.
- Set flow to 50 slpm and record reading. Report reading and uncertainty on the Data Reporting Sheet.
- Set flow to 75 slpm and record reading. Report reading and uncertainty on the Data Reporting Sheet.
- Set flow to 100 slpm and record reading. Report reading and uncertainty on the Data Reporting Sheet.
- Set flow to 50 slpm and record reading. Report reading and uncertainty on the Data Reporting Sheet.
- Email results and your uncertainty budget to proficiency@lakebalance.com
- After you've received your completed test results return to Lake Balance Calibration Solutions.
- Flow Meter Operating Instructions
- To set gas type:
- Press Menu (bottom right) and then Basic Config (bottom left).
- Press Gas (top left) scroll to recent and press Select (bottom right).
- Use Up and Down buttons (top left & center) to scroll to Recent and press Select (bottom right).
- Use Up and Down buttons (top left & center) to scroll to Air and press Select (bottom right).